USER'S GUIDE ____________________________________________________________________
42 ___________________________________________________________________ M211435EN-E
Configuring Analog Output
Set Analog Output Parameters and
Scaling
Use the ASEL command to show or set the output parameter and the
scaling of the analog output.
ASEL [parameter] [lowlimit highlimit]<cr>
where
parameter
=
Parameter that is output on analog channel.
Available parameters are TDF, TDFA, H2O
. If the
transmitter has been
ordered from Vaisala with a relative
humidity output
on the analog channel, also parameter RH
is available.
lowlimit
=
Lower limit of parameter scaling.
highlimit
=
Higher limit of parameter scaling.
Example (show current settings):
asel
Ch1 Tdf lo : -80.00 'C ?
Ch1 Tdf hi : 20.00 'C ?
Example (change output parameter to H2O, set low and high limits):
asel h2o 0 6000
Ch1 H2O lo : 0.00 ppm
Ch1 H2O hi : 6000.00 ppm