4. Unfasten the locking screws on the cooling bar so that you can
push the probe into the bar.
NOTE
Push the probe deep enough: make sure that the marking hole meets the
end of the bar.
5. Lock the probe in place by tightening the locking screws on the
bar.
Probe and Cooling Set Mounting with Process Shut Down
- Complete the entire mounting process as instructed on page 37.
Probe and Cooling Set Mounting with Process Running
- First mount the cooling element (and fins) to warm it up; tightly plug
the hole for the probe in the cooling element for the duration of the
warmup period.
- After a few hours, complete the installation by installing the probe.
This procedure should lessen the amount of condensation that builds up
in the hole.
Figure 83 Insulation with Mineral Wool