User's Guide ______________________________________________________________________
38 __________________________________________________________________ M211060EN-A
Set Probe Address
To operate in the POLL mode, the probe must have an address. If
multiple probes share the same serial line, each probe must have a
different address.
For a description of the serial interface modes, see section Set Serial Line
Settings on page 35.
ADDR [nn]<cr>
where
nn = address (0 ... 99)
Example:
addr
Address : 0
Set Serial Interface Delay
User the SDELAY command to view or set the serial interface answer
minimum delay.
SDELAY [delay]<cr>
where
delay = Range 0 ... 255. Value corresponds to four
milliseconds (for example, 5 = 0.020 second
minimum answer delay)
Example:
sdelay
Serial delay : 30
sdelay 50
Serial delay : 50