User's Guide _______________________________________________________________________
44 ____________________________________________________________________M210498en-A
1. When adjusting both relative humidity and temperature, please make
the temperature adjustment first.
2. Check that the MM70 is switched on.
3. Take out the screw from the probe handle to expose the calibration
button. Press the button with a small screwdriver. When pressing the
button, indicator turns to adjusting mode.
4. Press OK to start the adjustment mode.
5. Select RH, press SELECT.
6. Press YES to check the environmental settings, to continue directly
adjustment press
NO.
7. Now the adjustment mode is on, press ADJUST to select the
adjustment method.
8. Select 2-point adjustment, press SELECT. Press OK to continue.
9. Set the probe to a lower reference relative humidity. If using
HMK15 calibrator, use the adapter fitting (13.5 mm hole).
You can follow the stabilization from the GRAPH display. Press
READY when the reading is stabilized.
10. Give the lower reference humidity value by using the arrow buttons.
Press OK.
11. Set the probe to the higher reference humidity. Press READY when
the reading is stabilized. You can follow the stabilization from the
GRAPH display.
12. Give the higher reference relative humidity value by using the arrow
buttons. Press
OK.
13. Confirm the adjustment, press YES (by pressing NO you return
to adjustment mode display and no changes are made). If the
difference between two references is less than 50 %, adjustment can
not be done.
14. Adjustment has been carried out. Press BACK to exit the
adjustment mode press EXIT to return to the basic display.
15. Replace the screw onto calibration button.
You can attach a sticker on the calibration button screw to seal the
calibration.