Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 35
Output the Measurement Message Once
Use the SEND command to output the measurement values once. If the
probe is in POLL mode and the line is not open for commands, specify
the address of the probe to receive the measurement message.
SEND [aaa]<cr>
where
aaa = Address of the probe, range 0 ... 254. Set with the ADDR
command.
Example (probe in STOP mode, no address needed):
send
T= 22.7 'C RH= 20.0 %RH Td= -1.5 'C
Example (probe in POLL mode, with address 10):
send 10
T= 22.8 'C RH= 20.1 %RH Td= -1.3 'C
Configuring Serial Line Operation
Set Serial Line Settings
Use the SERI command to show or set the serial line settings. The new
settings will be taken into use when the probe is reset or powered up.
SERI [b p d s]<cr>
where
b = baud rate
(300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600)
p = parity (n = none, e = even, o = odd)
d = data bits (7 or 8)
s = stop bits (1 or 2)
Example (shows default settings):
seri
Baud P D S : 19200 N 8 1