Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 117
Setting Relay Outputs
NOTE
When having only one relay module installed, its relays are called 'relay
1' and 'relay 2'.
When having two relay modules, the relays of the module connected to
slot MODULE 1 are called 'relay 1' and relay 2' and relays connected to
slot MODULE 2 are called 'relay 3' and 'relay 4'
0706-003
Figure 63 Relay Indicators on Display
The following number refers to Figure 63 above:
1 = Lists enabled relays. Activation state shown in black.
Disabled relays are not shown.
Use the display/keypad to set the relay outputs.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Interfaces, confirm by pressing the ►arrow button.
3. Select Relay outputs, confirm by pressing the ►arrow button.
4. Select Relay 1/2/3/4, confirm by pressing the ►arrow button.
5. Select the Quantity, confirm by pressing Change. Select the
Quantity by using the arrow buttons. Confirm your selection by
pressing Select.
6. Select Act. above / Act. below. Press SET to confirm your
selection. (If asked, select MODIFY if you want to set the setpoint
by using the arrow buttons. Select REMOVE if you want to
remove the setpoint.)
7. Select Hysteresis by using the arrow buttons. Press SET to set the
hysteresis. Press OK.
8. Select Relay enable by using the arrow buttons, press ON/OFF to
enable/disable the relay.