Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 33
Ch1 aw hi : 1.00 ?
Ch2 T lo : -40.00 'C ?
Ch2 T hi : 60.00 'C ?
>asel t ppm
Ch1 T lo : -40.00 'C ?
Ch1 T hi : 60.00 'C ?
Ch2 ppm lo : 0.00 ppm ?
Ch2 ppm hi : 5000.00 ppm ?
>
*
ppm calculation only for transformer oils
ASCL Scaling the analog outputs
ASCL
Example:
>ascl
Ch1 T lo : -40.00 'C ? -20
Ch1 T hi : 60.00 'C ? 40
Ch2 ppm lo : 0.00 ppm ? 0
Ch2 ppm hi : 5000.00 ppm ? 3000
>ascl
Ch1 T lo : -20.00 'C ?
Ch1 T hi : 40.00 'C ?
Ch2 ppm lo : 0.00 ppm ?
Ch2 ppm hi : 3000.00 ppm ?
*
ppm calculation only for transformer oils
ITEST Testing the analog outputs
The operation of the analog outputs are tested by forcing the outputs
to given values. The values in the analog outputs can then be
measured with a current/voltage meter.
ITEST aa.aaa bb.bbb
where: aa.aaa = current value to be set for channel 1
bb.bbb = current value to be set for channel 2
mA
Vedere la pagina 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 62 63

Commenti su questo manuale

Nessun commento