Diagnostic commands
The PTU200 series digital transmitters respond to a set of diagnostic
commands, which are useful for analyzing the basic settings and
general operation of the transmitters. The diagnostic commands are
listed in
TABLE 7-2 Diagnostic commands
Function Command
basic information on settings ?
software version VERS
serial number SNUM
outputting error messages ERRS
? Basic information on the transmitter settings
? <cr>
The ? command lists the basic information of the transmitter.
>?
Software version PTU200 / 1.02
Serial number T5030004
Configuration 12
Linear adjustments ON
Multipoint adjustments ON
Calibration date 1999-03-12
Baud Parity Data Stop Dpx 9600 E 7 1 F
Echo ON
Sending mode STOP
Address 0
Output interval 3 s
Output format 4.1P""UUU""2.1RH""UUU#r#n
Error output format
SCOM format
Pressure unit hPa
Temperature unit 'C
Transducer 1:
TYPE: PMT
Transducer 2:
TYPE: VMT
Transducer 3:
TYPE: NONE
>
NOTE
Use this command only for one transmitter at a time. Any PTU200
series transmitter will always respond to ?-command whatever its
settings are, provided that the serial interface settings are correct.