Chapter 9 ______________________________ Field Checking of Fixed Transmitters
VAISALA __________________________________________________________43
CHAPTER 9
FIELD CHECKING OF FIXED
TRANSMITTERS
This chapter describes the field checking of fixed transmitters.
Field Checking of the Vaisala GMW115 and
GMD/W20 Series Transmitters (Diffusion
Sampling)
0505-235
Figure 15 Location of the MI70 Connector Ports
The following number refers to Figure 15 above:
1 = Connectors for the GMA70 cable
Check the reading of a wall mount or a duct mount transmitter by using
the GM70 as a reference. Both readings can be seen on the GM70's
display simultaneously when using a connecting cable GMA70 (see
section Accessories on page 65).