E28...E31 Unknown/incompatible module
installed in add-on module slot 1/2/
3/4)
Ensure that the module is compatible with the
PTB330.
E4 Pressure out of valid range Check that the assumed pressure is within the
measurement range for the barometer.
E3 Difference between pressure
transducers too large
1) Check that the barometer modules are
measuring the same pressure or 2) Check if one
of the barometer modules is out of the valid range
or 3) Check if the DPMAX value is set too low.
E16...E19 Pressure measurement failure on
add-on module 1/2/3/4
Internal barometer failure. Return the barometer
to the Vaisala Service Center.
E9 Checksum error in the internal
configuration memory
Internal barometer failure. Return the barometer
to the Vaisala Service Center.
E24...E27 EEPROM failure on add-on
module 1/2/3/4
Internal barometer failure. Return the barometer
to the Vaisala Service Center.
Table 14 Error Messages
Error Code Error Message Action