6.1.3 Humidity calibration procedure after sensor change
Humidity calibration should be performed in stable conditions using saturated
salt solutions as a reference. If the transmitter has the re-gaining option, the
re-gaining has to be done manually before starting the calibration.
6.1.3.1 With serial commands
• Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the sensor head.
Disconnect the security lock jumper!
• Insert the sensor head into the measurement hole of the
LiCl chamber.
• Wait at least 20 minutes for the reading to stabilize.
• Give command FCRH <cr>, enter the first point value and
press <cr>:
>FCRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...
• The stabilization of the sensor to the reference humidity
can be monitored by entering c <cr>:
RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...
• Insert the sensor head into the measurement hole of the
NaCl chamber.
• Wait at least 20 minutes for the reading to stabilize.
• Press any key, enter the second point value and press <cr>.
RH : xx.x Ref2 ? yy.y <cr>
• The stabilization of the sensor can be monitored by
entering c <cr>.
6.1.3.2 With display / keypad commands
• Leave the calibrator and the transmitter for at least 4 hours
in the same space so that their temperatures have time to
equalize. Remove the filter cap on the sensor head.
Disconnect the security lock jumper!