Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 117
To configure the sensor heating use the XHEAT command without
parameters. Enter the values after question mark. The available ranges
include the following:
Extra heat RH -limit (heating
function starts-up above the
setpoint)
0...100 %RH (default: 95 %RH)
Extra heating temperature 0...200 ºC (default: 100 ºC)
Extra heating time 0...255 s (default: 30 s)
Example:
>xheat
Extra heat : OFF
Extra heat RH : 95 ? 90
Extra heat temp: 100 ? 85
Extra heat time: 30 ? 10
>xheat on
Extra heat : ON
>
Vedere la pagina 118