User's Guide ______________________________________________________________________
96 __________________________________________________________________ M210796EN-A
Example: 0 ... 5 V voltage output selected for channel 1 and 4...20 mA
selected for channel 2.
OFF ON Selection
1
2
Volta
g
e out
p
ut selecte
d
3
4
5
6
0...5 V selecte
d
7
8
1
Current out
p
ut selecte
d
2
3
4
4... 20 mA selecte
d
5
6
7
8
NOTE
If you have customized the error output setting (AERR), check that
the set error values are still valid after changing the output
mode/range, see section Analog Output Fault Indication Setting on
page 99.
Analog Output Quantities
NOTE
For best accuracy, pressure must always be output using Ch3, if
available.
Use the display/keypad to change and scale the analog output
quantities.
1. Press any of the arrow keys to open the MAIN MENU.
2. Select Interfaces and press the right arrow key.
3. Select Analog outputs and press the right arrow key.
4. Select Output 1/2/3 and press the right arrow key.
Vedere la pagina 97