Chapter 6 ___________________________________________________Calibration and adjustment
VAISALA ______________________________________________________________________ 139
Using Serial Line
CTEXT
Use the CTEXT command to enter text to the adjustment information
field.
Example:
>ctext
Adjust. info : (not set) ? HMK15
>
CDATE
Use the CDATE command to enter date to adjustment information
field. Set the adjustment date in format YYYY-MM-DD.
Example:
>cdate
Adjust. date : (not set) ? 2006-01-22
>